ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 154

148
zoll.com
9650-001358-01 Rev. B
IBP2 (cont’d)
P2 Diastolic Alarm
(cont’d)
Lower Limit
Adult: 35
Pediatric: 35
Neonate: 30
-30-298
P2 Mean Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 120
Pediatric: 110
Neonate: 80
-28-300
Lower Limit
Adult: 50
Pediatric: 50
Neonate: 35
-30-298
Source Label
P2
P2, ABP, AO, ART, BAP,
CVP, FAP, ICP, LAP, PAP,
RAP, UAP, UVP
Display Format
S/D (M)
S/D (M), S/D, (M) S/D, (M)
IBP3
P3 Systolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 220
Pediatric: 145
Neonate: 100
-28-300
Lower Limit
Adult: 75
Pediatric: 75
Neonate: 50
-30-298
P3 Diastolic Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 110
Pediatric:100
Neonate: 70
-28-300
Lower Limit
Adult: 35
Pediatric: 35
Neonate: 30
-30-298
P3 Mean Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 120
Pediatric: 110
Neonate: 80
-28-300
Lower Limit
Adult: 50
Pediatric: 50
Neonate: 35
-30-298
Menu
Submenu
Parameter Name
Default Value
Parameter Values