NIBP Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
27
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
NIBP MAP Alarm
The following options allow you to set the limits of the NIBP MAP alarm for all patient age
groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Possible Values:
Adult: 20-218
Pediatric: 20-128
Neonate: 10-108
Default:
Adult: 35
Pediatric: 35
Neonate: 30
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
Adult: 25-230
Pediatric: 25-140
Neonate: 15-110
Default:
Adult: 120
Pediatric: 110
Neonate: 80