NIBP Parameter Settings

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

27

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

NIBP MAP Alarm

The following options allow you to set the limits of the NIBP MAP alarm for all patient age
groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Possible Values:

Adult: 20-218

Pediatric: 20-128

Neonate: 10-108

Default:

Adult: 35

Pediatric: 35

Neonate: 30

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

Adult: 25-230

Pediatric: 25-140

Neonate: 15-110

Default:

Adult: 120

Pediatric: 110

Neonate: 80