ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 160

154
zoll.com
9650-001358-01 Rev. B
Alarms (cont’d)
Default
Pediatric>FiCO2
Alarm
Upper Limit Enable
Off
On, Off
Upper Limit
8
2-98
Default Pediatric>RR/
BR Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
50
5-150
Lower Limit
3
0-145
Default Pediatric>No
Breath Alarm
No Breath Alarm
Off
Off, 10 sec, 15 sec, 20 sec,
25 sec, 30 sec, 40 sec, 50
sec, 60 sec
Default
Pediatric>SpO2 Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
100
52-100
Lower Limit
85
50-98
Default
Pediatric>SpCO
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
10
2-100
Lower Limit
0
0-98
Default
Pediatric>SpMet
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
3
2.0-100.0
Lower Limit
0
0.0-98.0
Default Pediatric>T1
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
Default Pediatric>T2
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Menu
Submenu
Parameter Name
Default Value
Parameter Values