ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 157

Advertising
background image

Supervisor Menu

9650-001358-01 Rev. B

ZOLL Configuration Manual

151

Alarms (cont’d)

Default Adult>RR/BR
Alarm (cont’d)

Lower Limit Enable

Off

On, Off

Upper Limit

50

5-150

Lower Limit

3

0-145

Default Adult>No
Breath Alarm

No Breath Alarm

Off

Off, 10 sec, 15 sec, 20 sec,
25 sec, 30 sec, 40 sec, 50
sec, 60 sec

Default Adult>SpO2
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

100

52-100

Lower Limit

85

50-98

Default Adult>SpCO
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

10

2-100

Lower Limit

0

0-98

Default Adult>SpMet
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

3.0

2.0-100.0

Lower Limit

0.0

0.0-98.0

Default Adult>T1
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

°F: 100.0

°C: 37.8

°F: 34.0-122.0

°C: 2.0-50.0

Lower Limit

°F: 95.0

°C: 35.0

°F: 32.0-120.0

°C: 0.0-48.0

Default Adult>T2
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

°F: 100.0

°C: 37.8

°F: 34.0-122.0

°C: 2.0-50.0

Lower Limit

°F: 95.0

°C: 35.0

°F: 32.0-120.0

°C: 0.0-48.0

Default Adult>

T

Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Menu

Submenu

Parameter Name

Default Value

Parameter Values