SpO2 Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
33
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
SpCO Alarm
The following options allow you to set the limits of the SpCO alarm (if enabled) for all patient
age groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Lower Limit
The Lower Limit parameter allows you to set the lower limit
.
The values for the Lower Limit parameter are:
Possible Values:
50-98
Default:
85
Possible Values:
On, Off
Default:
On
Possible Values:
On, Off
Default:
Off
Possible Values:
2-100
Default:
Adult: 10
Pediatric: 10
Neonate: 10
Possible Values:
0-98
Default:
Adult: 0
Pediatric: 0
Neonate: 0