IBP2 Parameter Settings

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

45

P2 Mean Alarm

The following options allow you to set the limits of the P2 Mean alarm for all patient age
groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

Source Label

The Source Label parameter allows you to specify the label of the P2 channel.

The values for the Source Label parameter are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

-28-300

Default:

Adult: 120

Pediatric: 110

Neonate: 80

Possible Values:

-30-298

Default:

Adult: 50

Pediatric: 50

Neonate: 35

Possible Values:

P2, ABP, AO, BAP, CVP, FAP, ICP, LAP, PAP, RAP, UAP, UVP

Default:

P2