Nibp parameter settings, Nibp systolic alarm – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 31

NIBP Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
25
NIBP Parameter Settings
The NIBP parameter settings allow you to view and modify functions related to using the X
Series unit for NIBP monitoring, and includes:
•
NIBP Systolic Alarm
•
NIBP Diastolic Alarm
•
NIBP MAP Alarm
•
NIBP Mode
•
NIBP Auto Mode Interval
•
SmartCuf On/Off
NIBP Systolic Alarm
The following options allow you to set the limits of the NIBP Systolic Alarm for all patient age
groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
Adult: 32-260
Pediatric: 32-160
Neonate: 22-120
Default:
Adult: 220
Pediatric: 145
Neonate: 100