T alarm – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 68

Chapter 4
Supervisor Menu
62
zoll.com
9650-001358-01 Rev. B
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
T Alarm
The following options allow you to set the limits of the adult
T alarm
.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit parameter are:
Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
Possible Values:
°F: 32.0-120.0
°C: 0.0-48.0
Default:
°F: 95.0
°C: 35.0
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
°F: 0.2-90.0
°C: 0.2-50.0
Default:
°F: 5.0
°C: 2.8
Possible Values:
°F: 0.0-89.8
°C: 0.0-49.8
Default:
°F: 0.0
°C: 0.0