ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 151

Parameter Menus
9650-001358-01 Rev. B
ZOLL Configuration Manual
145
NIBP (cont’d)
NIBP MAP Alarms
(cont’d)
Lower Limit
Adult:
50
Pedi
atric: 35
Neonate: 35
Adult: 23-228
Pediatric: 23-138
Neonate: 13-108
NIBP Mode
Manual
Manual, Auto
NIBP Auto Mode
Interval
5 min
1 min, 2 min, 3 min, 5 min,
10 min, 15 min, 30 min, 60
min
SmartCuf On/Off
On
On, Off
Resp
RR/BR Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 50
Pediatric: 50
Neonate: 80
5-150
Lower Limit
Adult: 3
Pediatric:3
Neonate: 12
0-145
No Breath Alarm
Adult: Off
Pediatric: Off
Neonate: Off
Adult: Off, 10 sec, 15 sec, 20
sec, 25 sec, 30 sec, 40 sec,
50 sec, 60 sec
Pediatric: Off, 10 sec. 15
sec, 20 sec, 25 sec, 30 sec,
40 sec, 50 sec, 60 sec
Neonate: Off, 10 sec, 15 sec,
20 sec
Resp Monitoring
On
On, Off
Resp Lead
Lead II (RA-LL)
Lead I (RA-LA), Lead II (RA-
LL)
CO2/Resp Sweep
Speed
12.5 mm/s
3.13 mm/s, 6.25 mm/s, 12.5
mm/s
SpO2
SpO2 Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 100
Pediatric: 100
Neonate: 95
52-100
Lower Limit
85
50-98
SpCO Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
Adult: 10
Pediatric: 10
Neonate: 10
2-100
Menu
Submenu
Parameter Name
Default Value
Parameter Values