ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 151

Advertising
background image

Parameter Menus

9650-001358-01 Rev. B

ZOLL Configuration Manual

145

NIBP (cont’d)

NIBP MAP Alarms
(cont’d)

Lower Limit

Adult:

50

Pedi

atric: 35

Neonate: 35

Adult: 23-228

Pediatric: 23-138

Neonate: 13-108

NIBP Mode

Manual

Manual, Auto

NIBP Auto Mode
Interval

5 min

1 min, 2 min, 3 min, 5 min,
10 min, 15 min, 30 min, 60
min

SmartCuf On/Off

On

On, Off

Resp

RR/BR Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 50

Pediatric: 50

Neonate: 80

5-150

Lower Limit

Adult: 3

Pediatric:3

Neonate: 12

0-145

No Breath Alarm

Adult: Off

Pediatric: Off

Neonate: Off

Adult: Off, 10 sec, 15 sec, 20
sec, 25 sec, 30 sec, 40 sec,
50 sec, 60 sec

Pediatric: Off, 10 sec. 15
sec, 20 sec, 25 sec, 30 sec,
40 sec, 50 sec, 60 sec

Neonate: Off, 10 sec, 15 sec,
20 sec

Resp Monitoring

On

On, Off

Resp Lead

Lead II (RA-LL)

Lead I (RA-LA), Lead II (RA-
LL)

CO2/Resp Sweep
Speed

12.5 mm/s

3.13 mm/s, 6.25 mm/s, 12.5
mm/s

SpO2

SpO2 Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 100

Pediatric: 100

Neonate: 95

52-100

Lower Limit

85

50-98

SpCO Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 10

Pediatric: 10

Neonate: 10

2-100

Menu

Submenu

Parameter Name

Default Value

Parameter Values