CPR Parameters

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

133

Enable Metronome Set to Always

When Enable Metronome is set to Always and Constant Metronome is Enabled, the metronome
is on when CPR is detected, and off when CPR is idle in all modes.

When Enable Metronome is set to Always and Constant Metronome is Disabled:

Manual Mode and Advisory/CPR Protocol Mode -- Metronome is only on compressions
when compressions are less than 80 per minute during CPR.

AED Mode -- Metronome is on at the start of CPR period.

Enable Metronome Set to AED Mode Only

When Enable Metronome is set to AED Mode Only and Constant Metronome is Enabled:

Manual Mode Advisory/CPR Protocol Mode- No metronome.

AED Mode -- Metronome is on when CPR is detected, and off when CPR is idle.

When Enable Metronome is set to AED Mode Only and Constant Metronome is Disabled:

Manual Mode Advisory/CPR Protocol Mode-- No metronome.

AED Mode -- Metronome is on at the start of CPR period.

The values for the Enable Metronome parameter are:

Table 1: CPR Metronome Activity

Possible Values:

Never, Always, AED Mode Only

Default:

Never

Enable
Metronome
Setting

AED Mode

Advisory/CPR
Protocol Mode

Manual Mode

Constant Metronome Enabled

AED Mode
Only

On when CPR detected
Off when CPR Idle

Off

Off

Always

On when CPR detected
Off when CPR Idle

On when CPR detected
Off when CPR Idle

On when CPR detected
Off when CPR Idle

Never

Off

Off

Off

Constant Metronome Disabled

AED Mode
Only

On at start of CPR period

Off

Off

Always

On at start of CPR period

On only at <80
compressions per minute

On only at <80
compressions per minute

Never

Off

Off

Off