Chapter 3

Parameter Menus

36

zoll.com

9650-001358-01 Rev. B

Temp Parameter Settings

The Temp parameters allow you to view and modify functions related to using the X Series unit
for temperature monitoring. These parameters include:

T1 Alarm

T2 Alarm

T

Alarm

T1 Source Label

T2 Source Label

T1 Alarm

The following options allow you to set the limits of the T1 alarm for all patient age groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

°F: 34.0-122.0

°C: 2.0-50.0

Default:

°F: 100.0

°C: 37.8

Possible Values:

°F: 32.0-120.0

°C: 0.0-48.0

Default:

°F: 95.0

°C: 35.0