HR/PR Parameter Settings

The HR/PR parameter settings allow you to view and modify functions related to using the X
Series unit to monitor heart rate. These parameters include:

HR/PR Alarm

Extreme Brady/Tachy Alarms

LTA (Life Threatening Alarms) Monitoring

HR/PR Tone

HR/PR Selected Source

Current Source

ECG Sweep Speed

HR/PR Alarm

The following options allow you to set the limits of the heart rate/pulse rate alarm for all patient
age groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:.

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

32-300

Default:

Adult: 120

Pediatric: 150

Neonate: 200