HR/PR Parameter Settings

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

23

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

Extreme Brady/Tachy Alarms

If LTA monitoring is enabled, the Extreme Brady/Tachy Alarms parameter allows you to set
the limits of the Extreme Brady/Tachy alarms.

The values for the Extreme Brady/Tachy Alarm parameter are:

LTA Monitoring

The LTA Monitoring parameter allows you to enable or disable LTA monitoring.

The values for the LTA Monitoring parameter are:

HR/PR Tone

The HR/PR Tone parameter allows you to enable or disable the HR/PR tone.

The values for the HR/PR Tone parameter are:

HR/PR Selected Source

The HR/PR Selected Source parameter allows you to specify the HR/PR source.

The values for the HR/PR Selected Source parameter are:

Possible Values:

30-298

Default:

Adult: 50

Pediatric: 50

Neonate: 100

Possible Values:

Tachy: +5-+30 by 5’s

Brady: -5--30 by 5’s

Default:

Tachy: +20

Brady: -10

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

On

Possible Values:

ECG, SpO2, P1, P2, P3

Default:

ECG