Lower Limit
The Lower Limit parameter allows you to set the lower limit.
The values for the Lower Limit parameter are:
No Breath Alarm
The No Breath Alarm parameter allows you to set the time of the No Breath alarm for all
patient age groups.
The values for the No Breath Alarm parameter are:
Resp Monitoring
The Resp Monitoring parameter allows you to enable or disable Resp monitoring.
The values for the Resp Monitoring parameter are:
Resp Lead
The Resp Lead parameter allows you to specify the Lead cable used for Resp monitoring.
The values for the Resp Lead parameter are:
Possible Values:
0-145
Default:
Adult: 3
Pediatric:3
Neonate: 12
Possible Values:
Adult: Off, 10 sec, 15 sec, 20 sec, 25 sec, 30 sec, 40 sec, 50 sec, 60 sec
Pediatric: Off, 10 sec. 15 sec, 20 sec, 25 sec, 30 sec, 40 sec, 50 sec, 60
sec
Neonate: Off, 10 sec, 15 sec, 20 sec
Default:
Adult: Off
Pediatric: Off
Neonate: Off
Possible Values:
On, Off
Default:
On
Possible Values:
Lead I (RA-LA), Lead II (RA-LL)
Default:
Lead II (RA-LL)