Alarms parameters, General – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 58

Chapter 4
Supervisor Menu
52
zoll.com
9650-001358-01 Rev. B
Alarms Parameters
The Alarm parameters allow you to view and modify functions related with Alarm states and
thresholds for the X Series unit. The main level parameters in this menu include:
•
General
•
Default Adult
•
Default Pediatric
•
Default Neonate
General
This following options allow you to specify basic conditions for all alarms.
Default Patient Mode
The Default Patient Mode parameter allows you to specify the default age group of the patients
using the X Series unit.
The values for the Default Patient Mode parameter are:
Volume
The Volume parameter allows you to set the volume level for all alarms.
The values for the Volume parameter are:
Alarm Suspend Time
The Alarm Suspend Time parameter allows you to specify how long any alarm is suspended.
The values for the Alarm Suspend Time parameter are:
Possible Values:
Adult, Pediatric, Neonate
Default:
Adult
Possible Values:
Low, Medium, High
Default:
Medium
Possible Values:
Disable, 2 min, 4 min, 15 min, Always On
Default:
4 min