ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 163

Supervisor Menu
9650-001358-01 Rev. B
ZOLL Configuration Manual
157
Alarms (cont’d)
Default
Neonate>EtCO2
Alarm (cont’d)
Lower Limit Enable
Off
On, Off
Upper Limit
60
5-150
Lower Limit
8
0-145
Default
Neonate>FiCO2
Alarm
Upper Limit Enable
Off
On, Off
Upper Limit
8
2-98
Default Neonate>RR/
BR Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
80
5-150
Lower Limit
12
0-145
Default Neonate>No
Breath Alarm
No Breath Alarm
Off
Off, 10 sec, 15 sec, 20 sec
Default
Neonate>SpO2 Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
95
52-100
Lower Limit
85
50-98
Default
Neonate>SpCO Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
10
2-100
Lower Limit
0
0-98
Default
Neonate>SpMet
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
3
2.0-100.0
Lower Limit
0
0.0-98.0
Default Neonate>T1
Alarm
Upper Limit Enable
Off
On, Off
Lower Limit Enable
Off
On, Off
Upper Limit
°F: 100.0
°C: 37.8
°F: 34.0-122.0
°C: 2.0-50.0
Lower Limit
°F: 95.0
°C: 35.0
°F: 32.0-120.0
°C: 0.0-48.0
Default Neonate>T2
Alarm
Upper Limit Enable
Off
On, Off
Menu
Submenu
Parameter Name
Default Value
Parameter Values