IM MW100-17E

When Action Occurs on Match Time

Setting

SLp1,p2,p3,p4,p5,p6<terminator>

p1 Box number (01 to 30)
p2 Event type (MATCH_TIME)
p3 Match time number (1 to 3)
p4 Event detection method
EDGE

Edge start

LEVEL

Level start

p5 Action types
p6 Flag numbers (F01 to F60)

Query

SL[p1]?

Example

Set box number 7 to an action in which recording starts

at the time specified for match time 1.
SL07,MATCH_TIME,1,EDGE,MEMORY_START

Explanation The p6 setting is only valid when p5 is FLAG. For the

action types, see section 1.3.

When Action Occurs on USER Key

Setting

SLp1,p2,p3,p4,p5,p6<terminator>

p1 Box number (01 to 30)
p2 Event type (USER_KEY)
p3 Key number (1, 2)
p4 Event detection method
EDGE

Edge start

LEVEL

Level start

p5 Action types
p6 Flag numbers (F01 to F60)

Query

SL[p1]?

Example

Set box number 8 to an action in which the specified file

is saved upon pressing of USER key 1.
SL08,USER_KEY,1,EDGE,FILE_SAVE

Explanation The p6 setting is only valid when p5 is FLAG. For the

action types, see section 1.3.

SI

Timer Settings

Setting

When p2=OFF

SIp1,p2<terminator>
When p2=RELATIVE
SIp1,p2,p3,p4<terminator>
When p2=ABSOLUTE
SIp1,p2,p3,p4<terminator>

When Not Using the Timer

Setting

SIp1,p2<terminator>

p1 Timer number (1 to 6)
p2 Timer type (OFF)

Query

SI[p1]?

Example

Set timer number 1 to OFF.

SI1,OFF

When Using the Relative Timer

Setting

SIp1,p2,p3,p4<terminator>

p1 Timer number (1 to 6)
p2 Timer type (RELATIVE)
p3 Time interval-days (0 to 31)

p4 Time interval-time (HH:MM)
HH hours (00 to 23)
MM

Minute (00 to 59)

(However, when p3 is 0, 00:00 cannot
be set.)

Query

SI[p1]?

Example

Set a relative time of 3 hours and 30 minutes on timer
number 1.
SI1,RELATIVE,0,03:30

When Using the Absolute Timer

Setting

SIp1,p2,p3,p4<terminator>
p1 Timer number (1 to 6)
p2 Timer type (ABSOLUTE)
p3 Time interval
1M/2M/3M/4M/5M/6M/10M/12M/15M/20M/30M

(minutes)

1H/2H/3H/4H/6H/8H/12H/24H (hours)
p4 Reference time (HH:MM)
HH hours (00 to 23)
MM

Minute (00 to 59)

Query

SI[p1]?

Example

Set a time interval of 30 minutes and a reference time of
17 hours and 15 minutes on timer number 1.
SI1,ABSOLUTE,30M,17:15

SQ

Match Time Settings

Setting

When p2=OFF
SQp1,p2<terminator>
When p2=MONTH
SQp1,p2,p3,p4<terminator>
When p2=WEEK
SQp1,p2,p3,p4<terminator>
When p2=DAY
SQp1,p2,p3<terminator>

When Not Using the Match Time

Setting

SQp1,p2<terminator>
p1 Match time number (1 to 3)
p2 Start type (OFF)

Query

SQ[p1]?

Example

Set match timer number 1 to OFF.
SQ1,OFF

When Starting Match Time in Units of Months

Setting

SQp1,p2,p3,p4<terminator>
p1 Match time number (1 to 3)
p2 Start type (MONTH)
p3 Start date/time-day (1 to 31)
p4 Start date/time-time (HH:MM)
HH hours (00 to 23)
MM

Minute (00 to 59)

Query

SQ[p1]?

Example

Set match time number 1 to 1 day, 8 hours 30 minutes.
SQ1,MONTH,1,08:30

1.4 Setting Commands (System Related)