Commands
1
2
3
App
App
Index
p2 Data type
MODBUS_MC Modbus master command
status
MODBUS_MC Modbus master connection
status
MODBUS_CC Modbus client command
status
MODBUS_CS Modbus client connection
status
BALANCE
Strain input initial
balance
p3 First data number
MODBUS_MC 1 to 100
MODBUS_MS 1 to 247
MODBUS_CC 1 to 100
MODBUS_CS 1 to 10
BALANCE
1 to 60
p4 Last data number
MODBUS_MC 1 to 100
MODBUS_MS 1 to 247
MODBUS_CC 1 to 100
MODBUS_CS 1 to 10
BALANCE
1 to 60
Example
Output the Modbus master connection status of data
number 1 to 10.
FL1,MODBUS_MS,1,10
IS
Output of Status Information
Setting
ISp1<terminator>
p1 Outputs status information (0)
Example
Output status information.
IS0
VF
Output of Relay Status
Setting
VFp1<terminator>
p1 Outputs relay status (0)
Example
Output relay status.
VF0
CF
Output of System Recognition
Status
Setting
CFp1<terminator>
p1 Outputs system recognition status (0)
Example
Output the system recognition status
CF0
ME
Output of Media Information
Setting
MEp1,p2<terminator>
p1 Output type
0
Outputs media free space
1
Outputs file information
p2 User name (up to 63 characters)
Enter using the full path.
Example
• Output file information.
ME0
• Output the information of MW100.txt in the DATA0
directory.
ME1,/DATA0/MW100.txt
• Output root directory information.
ME1,/
1.8 Output Commands