1.7 Control Commands/1.8 Output Commands/1.8 Output Commands

CS

Checksum Settings

Setting

CSp1<terminator>
p1 Checksum ON/OFF
0

None (do not calculate sum)

1

ON (calculate sum)

Query

CS?

Example

Turn checksum ON.
CS1

GR

Resetting of the MATH Channels

Specified by the MATH Group

Setting

GRp1<terminator>
p1 MATH group number
1 to 7 Reset the MATH channels of

the specified MATH group

Example

Reset the MATH channels of MATH group 1.
GR1

Explanation • You can set this command on products with the

/M1 MATH function option, or when the pulse input
module is installed.

• Resets the MATH channels assigned to a MATH

group using the MG command.

1.8 Output Commands

FD

Output of Latest Measured/

Computed Data

Setting

FDp1,p2,p3<terminator>
p1 Output data type
0

Outputs the latest measured/
computed data in ASCII

1

Outputs the latest measured/
computed data in BINARY

p2 First channel (001 to 060 and A001 to

A300)

p3 Last channel (001 to 060 and A001 to

A300)

Example

Output the most recent measured/computed data from
the instrument on channels 1 to 5 in ASCII format.
FD0,001,005

Explanation • The “latest measured/computed data,” refers to the

data most recently received into in the MW100’s
internal memory.

• If the first channel is 001 and the last channel is

A300, the data output is in the range 001 to 060 and
A001 to A300.

• Set a first and last channel so that the first is less

than or equal to the last.

FO

Output of Latest Output Data

Setting

FOp1,p2,p3<terminator>
p1 Output data type
0

Outputs the latest output
channel data in ASCII

1

Outputs the latest output
channel data in BINARY

p2 First channel (001 to 060)
p3 Last channel (001 to 060)

Example

Output the most recent output data from the instrument
on channels 1 to 5 in ASCII format.
FO0,001,005

Explanation Set a first and last channel so that the first is less than

or equal to the last.

FE

Output of Decimal Place, Units,

and Setting Data

Setting

FEp1,p2,p3<terminator>
p1 Output type
0

Outputs the setting command
query

1

Outputs the decimal place and
units

p2 First channel (001 to 060 and A001 to

A300)

p3 Last channel (001 to 060 and A001 to

A300)