Commands

1

2

3

App

App

Index

STOP2

STOP key (equivalent to
holding down for 2 sec)

USER1

USER1 key

USER2

USER2 key

Example

Executes the START key.
KISTART1

CE

Error Display Clear

Setting

CEp1<terminator>
p1 Clears the error display (0)

Example

Clears the error on the 7SEG display.
CE0

AK

Alarm Acknowledgment

Setting

AKp1<terminator>
p1 Check alarm status, and make

notification (0) of output clear
command for alarm relay output hold
status

Example

Check alarm status, and make notification of output
clear command for alarm relay output hold status.
AK0

IR

Timer Reset

Setting

IRp1<terminator>
p1 Number of timer
0

Reset all timers

1-6 Reset the specified timer

Example

Reset all timers.
IR0

CM

Communication Input Data

Settings

Setting

CMp1,p2<terminator>
p1 Communication input channel range

(C001 to C300)

p2 Communication input value
+INT (-∞)
-INF (+∞)
NAN (Not a Number)
-9.9999E+29 to -1.0000E-30
0
1.0000E-30 to 9.9999E+29

Query

CM[p1]?

Example

Set a communication input value of 1.0000E-10 to
communication input channel C001.
CMC001,1.0000E-10

Explanation See section 1.3 for the communication input value.

VD

Relay ON/OFF

Setting

VDp1,p2<terminator>
p1 Relay range (001 to 060)
p2 Relay status (OFF/ON)

Example

Turns the channel 1 relay ON.
VD001,ON

Explanation Valid only for the relay whose relay output type is set to

manual DO (COM) with the XD command.

SD

Date and Time Settings

Setting

SDp1,p2,p3<terminator>
p1 Date (YY/MM/DD)
YY

Year (80 to 99, 00 to 35)

MM

Month (01 to 12)

DD

Day (01 to 31)

p2 Time (HH:MM:SS)
HH hours (00 to 23)
MM

Minute (00 to 59)

SS

Seconds (00 to 59)

p3 Timezone (SHH:MM)
S

Symbol (+/-)

HH hours (00 to 23)
MM

Minute (00 to 59)

Query

SD?

Example

Sets the MW100’s internal time to April 1, 2005,
13:00:00. The time zone is 9:00.
SD05/04/01,13:00:00,+09:00

Explanation The positive time zone symbol can be omitted.

RC

Initialization of Settings

Setting

RCp1<terminator>
p1 Initialization level
0

Initialize all settings except
the module recognition

1

Initialize settings excluding
the following items

· IP address, host name, subnet

mask, default gateway

· Items relating to DHCP

· Items relating to DNS

· Items relating to

communication timeout

· Items relating to login

function

· Baud rate, parity bit, stop

bit, data length, handshaking,
and module recognition
information

Example

Initialize all settings.
RC0

Explanation The communication settings are the parameters set

using YS, YA, YD, YN, YH, UA, and YQ.

1.7 Control Commands