Rs-422a/485 specific commands/responses, Rs-422a/485 specific commands/responses -4 – Yokogawa Standalone MW100 User Manual

Page 54

Advertising
background image

2-4

IM MW100-17E

RS-422A/485 Specific Commands/Responses

Commands specific to the RS-422A/485 interface and their responses are shown in the
table below.

Command Syntax

Command Meaning Responses

ESC Oxx CRLF

Instrument open

• Response from instrument address specified

by command: ESC O xx CRLF

• No response when instrument address

specified by command does not exist *

ESC Cxx CRLF

Instrument close

• Response from instrument address specified

bycommand: ESC C xx CRLF

• No response when instrument address

specified by command does not exist *

* The reasons that an instrument of the specified address might not exist include a command

error, the specification does not match the address set on the instrument, the power to the

instrument is not turned ON, or the instrument is not connected by serial interface.

• The xx in the table indicates the address of the instrument. An address from 01 to

32 can be set, in the range of addresses set on the target instrument.

• Only one instrument can be open at a time.
• If an instrument is opened using the ESC 0 command, commands and responses

with that instrument are enabled.

• If one instrument is opened using the ESC 0 command, any other instrument that is

already opened is automatically closed.

• Most commands can use CR+LF or LF as a terminator, but these commands can

only use CF+LF.

Note

ESC is expressed as 0x1b in ASCII code. See appendix 5.

2.1 Response Format

Advertising