Rs reconfiguration, Mf media preparations, Bl execution of strain input initial balancing – Yokogawa Standalone MW100 User Manual

Page 47: Pv transmission output control, Sp user output settings, Ec execution of communication, Tc communication test, Cc communication open/close, Bo byte output order settings

Advertising
background image

1-40

IM MW100-17E

RS

Reconfiguration

Setting

RSp1<terminator>
p1 System reconfiguration (0)

Example

Perform system reconfiguration (re-detect the modules).
RS0

MF

Media Preparations

Setting

MFp1<terminator>
p1 Media initialization (0)

Example

Initialize external media (CF card).
MF0

BL

Execution of Strain Input Initial

Balancing

Setting

BLp1<terminator>
p1 Operation
0

Reset initial balance value

1

Execute initial balancing

Example

Executes initial balancing.
BL1

Explanation Valid only for channels on strain input modules on which

initial balancing is enabled (ON) with the BA command.

PV

Transmission Output Control

Setting

PVp1,p2<terminator>
p1 Channel range (001 to 060)
p2 Transmission output OFF/ON
OFF No transmission output
ON

Perform transmission output

Example

Turn OFF transmission output on channels 1-8.
PV001-008,OFF

Explanation Valid only for channels whose output method is set to

transmission output (TRANS) with the AO command.

SP

User Output Settings

Setting

SPp1,p2<terminator>
p1 Channel range (001 to 060)
p2 User output value (within the output
range’s upper and lower limits of span)

Example

Sets the user output value on channel 1 to 3 V.
SP001,3000

Explanation • Set the output range using the AO command.

• Valid only for channels whose output method is set

to arbitrary output (COM) with the AO command.

• Enter setting parameters for p2 excluding the

decimal place.

EC

Execution of Communication

Setting

ECp1<terminator>
p1 Operation
0

Gets the time (queries the time
on the SNTP server)

1.7 Control Commands/1.8 Output Commands

Example

Get the time.
EC0

Explanation Valid if the SNTP client function is ON.

TC

Communication Test

Setting

TCp1<terminator>
p1 Operation
FTP_C

FTP client test

SMTP

SMTP client test

p2 Connection (1, 2)

Example

Execute the FTP client test.
TCFTP_C,1

Explanation • The FTP client test is valid if the FTP client function

is ON.

• The SMTP client test is valid if the SMTP client

function is ON.

CC

Communication Open/Close

Setting

When p1=0
CCp1<terminator>
When p1=1
CCp1,p2,p3<terminator>

When Closing the Connection

Setting

CCp1<terminator>
p1 Closes the connection (0)

Example

Close the connection.
CC0

When Opening the Connection

Setting

CCp1,p2,p3<terminator>
p1 Opens the connection (1)
p2 User name (up to 16 characters)
p3 Password (max 8 characters)

Example

Try to open a connection using user name "ABC" and
password "123456."
CC1,ABC,123456

Explanation • For the characters that can be used for user names,

see user specified characters in section 1.3.

• For the characters that can be used for passwords,

see password strings in section 1.3.

BO

Byte Output Order Settings

Setting

BOp1<terminator>
p1 Byte output order
0

Outputs in order from MSB (most
significant byte)

1

Outputs in order from LSB (least
significant byte)

Query

BO?

Example

Output from the top byte.
BO0

Explanation This command operates per the byte order of the

numerical data during BINARY output.

Advertising