App index, Xv measurement interval settings, Vm math interval settings – Yokogawa Standalone MW100 User Manual

Page 32: Xg settings of operation upon math error

Advertising
background image

1-25

IM MW100-17E

Commands

1

2

3

App

App

Index

When Starting Match Time in Units of Weeks

Setting

SQp1,p2,p3,p4<terminator>
p1 Match time number (1 to 3)
p2 Start type (WEEK)
p3 Start Date/time-day of week
SUN Sunday
MON Monday
TUE Tuesday
WED Wednesday
THU Thursday
FRI Friday
SAT Saturday
p4 Start date/time-time (HH:MM)
HH hours (00 to 23)
MM

Minute (00 to 59)

Query

SQ[p1]?

Example

Set match time number 1 to Friday, 17 hours 30
minutes.
SQ1,WEEK,FRI,17:30

When Starting Match Time in Units of One Day

Setting

SQp1,p2,p3<terminator>
p1 Match time number (1 to 3)
p2 Start type (DAY)
p3 Start date/time-time (HH:MM)
HH hours (00 to 23)
MM

Minute (00 to 59)

Query

SQ[p1]?

Example

Set match time number 1 to 17 hours 30 minutes.
SQ1,DAY,17:30

XV

Measurement Interval Settings

Setting

XVp1,p2<terminator>
p1 Measurement group number (1 to 3)
p2 Measurement interval
OFF
10mS/50mS/100mS/200mS/500mS

(miliseconds)

1S/2S/5S/10S/20S/30S/60S (seconds)

Query

XV[p1]?

Example

Set measurement group number 1 to a measurement
interval of 1 second.
XV1,1S

XI

Measurement Groups and A/D

Integration Time Settings

Setting

XIp1,p2,p3<terminator>
p1 Module number (0 to 5)
p2 Measurement group number (OFF, 1 to

3)

p3 A/D integration time

(AUTO, 50 Hz, 60 Hz)

Query

XI[p1]?

Example

Set module number 0 to measurement group 1 and A/D
integration time to automatic detection.

XI0,1,AUTO

Explanation • If the module specified in p1 has a shortest

measurement interval of 100 ms, you cannot set
measurement group numbers having measurement
intervals of 10 ms or 50 ms. Also, if the module has
a shortest measurement interval of 500 ms, you
cannot set measurement group numbers having
measurement intervals of 10 ms, 50 ms, or 100 ms.

• Measurement group numbers cannot be specified for

output modules and unmounted modules.

• For the 30-CH Medium Speed DCV/TC/DI Input

Module, enter the same settings for the three
measurement groups and A/D integral times.

VM

MATH Interval Settings

Setting

VMp1<terminator>
p1 Measurement group number

(OFF, 1 to 3)

Query

VM?

Example

Set MATH interval to measurement group number 1.

VM1

Explanation • You can set this command on products with the /M1

MATH function option, and when the pulse input
module is installed.

• Measurement group numbers of measurement

interval 10 ms or 50 ms cannot be set.

XG

Settings of Operation upon

MATH Error

Setting

XGp1,p2,p3,p4,p5,p6,p7<terminator>
p1 Operation upon Math Error
+OVER

Set computed value to +Over

-OVER

Set computed value to -Over

p2 Special MATH processing when the

input value of the measurement input
channel is in error in a computation
other than TLOG and CLOG.

ERROR

Sets the computed value to
MATH error.

SKIP

Continue the computation
using the previous value for
channels with an error in
the input value.

p3 Special MATH processing when the

input value of the measurement input
channel is overflow in a computation
other than TLOG and CLOG.

ERROR

Compute using the overflow
value

SKIP

Continue the computation using
the previous value for channels
with an overflow value

1.4 Setting Commands (System Related)