Application Notes

SYS Mode

There are two methods for configuring a temperature to
electrical calibration test using SYS mode. The
following table indicates which method to use.

Use this method if…

Method 1

 Your field instrument has an

electrical output range within the
ability of the DC module on the
BetaGauge II.

Method 2

 Your field instrument has an

electrical output range that exceeds
the ability of the DC module on the
BetaGauge II.

NOTE: The BetaGauge II cannot make direct

temperature measurements.