Basic Operations

3-14

3.

Press the function key corresponding to the
damping setting you want to use, then press the
CLR key to return to the Setup menu.

CAUTION: Enabling the damping feature affects

your ability to accurately zero a module’s
displayed value. Therefore, prior to
zeroing the displayed value of a module,
disable the damping feature by setting the
damping value to 0.0.

Viewing Module Information

The BetaGauge II allows you to view the following
information for the modules currently installed in the
calibrator:

 Serial number

 Full scale range (pressure modules only)
 Last calibration date

Procedure

To view module information, do the following:

1.

Use the Menu key to locate the Setup menu item
Mod, then press the corresponding function key to
display the Module Info menu.

2.

Press the function key corresponding to the channel
you want to view. Remember that CH1 represents
the left module and CH2 represents the right
module. The lower portion of the LCD will appear
similar to the examples shown below:

DC Module Example:

DC SN: 610006

Cal: 09/25/2005