Martel Electronics BetaGauge 3II User Manual

Page 135

Advertising
background image

Application Notes

6-50

SYS Mode

There are two methods for configuring a pressure
indicator or loop calibration using SYS mode. The
following table indicates which method to use.

Use this method if…

Method 1

 Your field instrument has a pressure

input range within the ability of the
pressure module on the BetaGauge
II, and

 You are using either a local or

remote pressure indicator to indicate
the pressure input signal to the field
instrument.

Method 2

 Your field instrument has a pressure

input range that exceeds the ability
of the pressure module on the
BetaGauge II, and

 You are using either a local or

remote pressure indicator to measure
the pressure output signal from the
field instrument.

Method 1

Software Instrument Setup

 Pressure input
 Pressure indicator output (requires manual

entry on the BetaGauge II)

Other Necessary Equipment

 Pressure source