Accessing the execute level – Martel Electronics BetaGauge 3II User Manual

Page 154

Advertising
background image

7-13

Doc. Cal. Mode

Accessing the Execute Level

To access the Execute level of the SYS mode, press the
function key corresponding to Exe on the Back screen.
The Test Setup screen appears as shown below:

The Test Setup screen is described below and on the
following page.

Test Setup Screen Description

The main purpose of the Test Setup screen is to indicate
whether you have installed modules with the proper
range for the assigned test point values of the current
tag. In addition, you can ensure that the currently
installed modules have not exceeded their calibration
interval.

NOTE: You cannot proceed from the Test Setup

screen until you have installed modules that
will accommodate the assigned test point
values for the current tag.

Tag 2:
BG2 TAGS Tag 2

Inst. OUT:

mA

4.0000

20.000

CH1

Inst. IN:

PSI

3.0000

15.000

CH2

CH1

DC

09/25/1995
CH2

GAGE

04/16/1995

15.00

Modules Correct

Ready