Advanced Display Modes

5-4

Programming Display Modes for
BetaPort–DC Modules

The BetaGauge II allows you to define process signals
for display using any of the following modes:
 engineering units
 percent scale
 percent error
 square root extraction

NOTE: Before performing any of the procedures for

setting up display modes for BetaPort–DC
modules, ensure that you are familiar with the
input and output signal ranges of the field
instrument you are calibrating.

The following paragraphs describe how to set up the
display modes listed above.

Engineering Unit Display

When you want to display electrical signals in
engineering units, you must perform the following
procedures:

 Define a unit label, or name.
 Define the type of field instrument you want to

calibrate:
- Pressure transmitter (P/I)
- Electrical to pressure converter (I/P)
- Other

 Define the engineering low and high values with

respect to the expected field instrument output
signals.