Advanced Display Modes

5-8

5.

After you have specified all of the values in step 4,
the display screen prompts you for the number of
decimal places you want to display. A sample
screen is shown below:

6.

Use the keypad numbers to enter the number of
decimal places you want to appear on the display
(from 1 to 3), then press the ENTER key to return
to the Calib Instr menu.

7.

Press the CLR key to return to the Setup menu.
The BetaGauge II should now display the input
electrical signals using the values defined for the
engineering range.

Percent Scale Display

The BetaGauge II will display either or both of a field
instrument's input or output values in percent of scale.

When you want to display electrical signals in percent
of scale, you must perform the following procedures:

 Define the type of field instrument you want to

calibrate:
- Pressure transmitter (P/I)
- Electrical to pressure converter (I/P)
- Other

 Define values equal to the expected low and high

electrical signal values of the field instrument.

CHx Ent. Decimals: