Accuracy – Martel Electronics BetaGauge 3II User Manual

Page 179

Advertising
background image

Specifications

B-4

Accuracy

Pressure

Range

Accuracy

Overpressure

Notes

Gauge (PSIG)
0 to 0.3 (0 to 20mBar)

±0.1%

400 %

0 to 1 (0 to 67mBar)

±0.05%

400 %

0 to 5 (0 to 350 mBar)

±0.025%

400 %

9

0 to 7.2 (0 to 500 mBar)

±0.035%

300 %

4 & 7

0 to 10 (0 to 700 mBar)

±0.025%

300 %

9

0 to 30 (0 to 2 Bar)

±0.025%

300 %

0 to 50 (0 to 3.5 Bar)

±0.03%

300 %

0 to 100 (0 to 7 Bar)

±0.025%

300 %

0 to 150 (0 to 10 Bar)

±0.035%

200 %

4

0 to 300 (0 to 20 Bar)

±0.025%

200 %

0 to 500 (0 to 34 Bar)

±0.05%

200 %

0 to 1000 (0 to 70 Bar)

±0.05%

200 %

6

0 to 1500 (0 to 100 Bar)

±0.05%

200 %

4

0 to 3000 (0 to 200 Bar)

±0.05%

200 %

0 to 5000 (0 to 340 Bar)

±0.05%

200 %

0 to 10000 (0 to 700 Bar) ±0.05%

120 %

Vacuum (PSIG)
0 to -5 (0 to -350 mBar)

±0.025%

400 %

9

0 to -15 (0 to -1 mBar)

±0.025%

300 %

7

Absolute (PSIA)
0 to 15 (0 to 1 Bar)

±0.025%

300 %

7

0 to 30 (0 to 2 Bar)

±0.025%

300 %

0 to 50 (0 to 3.5 Bar)

±0.03%

300 %

0 to 100 (0 to 7 Bar)

±0.025%

300 %

0 to 300 (0 to 20 Bar)

±0.025%

200 %

Compound (PSIG)
-15 to 15 (-1 to 1 Bar)

±0.025%

300 %

7

-15 to 30 (-1 to 2 Bar)

±0.025%

300 %

Differential (PSID) (Note1)

0 to 5 (0 to 350 mBar)

±0.025%

400 %

5 & 8

0 to 30 (0 to 2 Bar)

±0.025%

300 %

5

0 to 50 (0 to 3.5 Bar)

±0.03%

300 %

5

Advertising