Input Modules

2-4

Procedure

Use the following procedure to connect the BetaPort–
DC electrical module to a field device:

NOTE: Install the module in the calibrator prior to

connecting the module to field devices.

1.

Insert the common pin (marked by a tab) on the
dual banana connector into the negative (black)
terminal marked with a "–" on the module.

2.

Insert the opposite pin into the positive (red)
terminal marked with a "+" on the side labeled "V"
for voltage measurement and "mA" for milliamp
measurement.

NOTE: The BetaPort– DC module is protected from

damage against misconnection to high voltages
up to 120 Volts. In the event that the module is
misconnected to a high voltage, the module
should be checked for correct operation by
comparison with another device of known
performance and similar or better accuracy.