Basic Operations

NOTE: The last selection in the list of pressure
units is User, which allows you to define your own
pressure unit for display purposes. Refer to Section
5 for more information.

4.

To return to the Setup menu, press the CLR key.

Selecting a Water Column Measurement

In the list of available display units for a BetaPort–P
pressure module, InWC and cmWC are water column
equivalent measurements. If you select a water column
pressure unit, you must also select the appropriate
temperature for water density correction. The available
temperature selections and their recommended use are
listed below:

Temperature

Recommended Use

4°C

Scientific applications

20°C

Industrial applications (ISA)

60°F

American Gas Association

If there is any question regarding temperature
compensation, consult the proper engineering authority
at your facility.