Basic Operations

3-16

Dry Contacts

To detect dry contact trips, you must connect dry
contacts to the milliamp input of the calibrator’s DC
module. The BetaGauge II will detect a dry contact trip
when the input resistance crosses the 800 Ohm
threshold.

Hints for Accurate Trip Measurements

There are three areas where it is helpful to have detailed
knowledge before setting up a pressure trip detection:

 Selecting a field instrument (hand pump or pressure

controller)

 Adjusting the pressure source
 Ramping the pressure signal

Field Instrument Selection

You can use either a hand pump or a pressure controller
as a pressure source for detecting pressure trips. Use the
following guidelines when selecting either of these field
instruments:
Hand Pump. Select a hand pump with a Vernier

adjustment knob so that you can make fine
adjustments.

Pressure Controller. Select a pressure controller

that has an easy micro-adjustment method.