Martel Electronics BetaGauge 3II User Manual

Page 97

Advertising
background image

Application Notes

6-12

SYS Mode

There are four methods for configuring a pressure to
electrical calibration test using SYS mode. The
following table indicates which method to use.

Use this method if…

Method 1

 Your field instrument has a pressure

output range within the ability of the
pressure module on the BetaGauge
II, and

 Your field instrument has an

electrical output range within the
ability of the DC module on the
BetaGauge II.

Method 2

 Your field instrument has a pressure

input range that exceeds the ability
of the pressure module on the
BetaGauge II, and

 Your field instrument has an

electrical output range within the
ability of the DC module on the
BetaGauge II.

Method 3

 Your field instrument has a pressure

input range within the ability of the
pressure module on the BetaGauge
II, and

 Your field instrument has an

electrical output range that exceeds
the ability of the DC module on the
BetaGauge II.