5-3

Advanced Display Modes

4.

When the unit label is satisfactory, press the
ENTER key once again to display the following
screen:

5.

Use the numbers on the keypad to specify the
1 PSI equivalent value of your user unit, then press
the ENTER key to return to the Setup menu.

NOTE: For the example Feet of Water Column at
20°C
, you would enter 2.3108.

6.

Use the MENU key to locate the Setup menu item
Unit, then press the corresponding function key to
display a screen similar to the following:

7.

Press the function key corresponding to → through
as many screens as necessary until the unit you
defined appears in the display.

8.

Press the function key associated with the pressure
unit you defined in step 3 and the pressure will
begin to display as specified.

NOTE: In the case of the example, the pressure
will begin to display in Feet of Water Column at
20°C
.

9.

Press the CLR key to return to the Setup menu.

Ent. 1PSI =

CHx

mmHg cmWC Kpa

→

Advertising