Martel Electronics BetaGauge 3II User Manual

Page 83

Advertising
background image

Advanced Display Modes

5-16

3.

Specify a unit label by using the CLR key to erase
the character (if any) to the left of the cursor and
use the arrow keys on the keypad to scroll through
the available character set. When the desired
character is displayed, press the ENTER key to
accept the character and move the cursor to the
right. A maximum of 4 characters may be entered
for identification.

NOTE: Pressing the ENTER key twice in a row
ends editing. Therefore, to specify the same
character consecutively, use the up and down arrow
keys to select the second character.

4.

When you have completed entering the unit label,
press the ENTER key once again to return to the
Setup menu.

Defining Output Milliamp Signal Values

1.

Use the Menu key to locate the Setup menu item
Mode, then press the corresponding function key to
display the following screen:

2.

Press the function key corresponding to the P/I
(pressure transmitter) menu item and the following
main linearization screen displays:

In the above example:
x = the channel number of the DC module,
yyyy = the current linearization mode of the
current DC module channel.

Calib Instr

P/I

I/P

Other

CHx Mode: yyyy

Std

Eng

Sqrt