Specifications

B-6

Environmental

Temperature

-20ºC to 50ºC (operating)

0ºC to 35ºC (storage with battery packs)

-30ºC to 70ºC (storage without battery packs)

Relative Humidity

0 to 95% non-condensing

Effects

Pressure: ± 0.0015% Range/ºC outside 15ºC to 35ºC

Electrical: ±0.002% Range/ ºC outside 15ºC to 35ºC

Overpressure

Non-isolated Sensors

0 to 5 psi

400% of sensor range

>5 psi

300% of sensor range

150 psi

200% of sensor range

Isolated Sensors

200% of sensor range

Alarm Trip Detect

Wet contact input

12 Vdc ± 10% threshold, 48 Vdc maximum

Dry contact input

800 Ohm ± 200 Ohm threshold,

4 mA maximum excitation current

Damping Factor

Programmable 2, 5, or 10 seconds response time for all
inputs (running average)