Page 95
Advertising
Application Notes
6-10
Pressure to Electrical Test
You can perform a pressure to electrical calibration test
using either the conventional mode or the SYS mode of
operation. The following paragraphs describe all of the
possible configurations, as well as, the other equipment
necessary to perform a pressure to electrical test.
Conventional Mode
You may perform this test if:
Your field instrument has a pressure input
range within the ability of the pressure module
on the BetaGauge II, and
Your field instrument has an electrical output
range within the ability of the DC module on
the BetaGauge II.
Field Instrument Setup
Pressure input
Electrical output
Other Necessary Equipment
Pressure source