7-23

Doc. Cal. Mode

Specifying Post-Test Status Information

NOTE: You must answer the post-test status prompts.

If you press the CLR key prior to answering
the post-test prompts, all calibration results
will be lost.

To specify post-test status information, do the
following;

1.

From the Post-Test Status screen, use the function
button corresponding to the appropriate menu item
(Yes, No, or Rep) to indicate whether the field
instrument was adjusted, left alone, or replaced.

2.

After you have specified the field instrument
adjustment status information, the BetaGauge II
prompts you for an identification number of the
technician who performed the calibration test.

3.

Do one of the following:

 Use the numbers on the keypad to specify a

technician’s identification number, then press
the ENTER key.

 Accept the 0 default value and press the

ENTER key to bypass the entry.

4.

After you have specified the identification number,
or bypassed the prompt, the display returns to the
Tag Status screen of the Main level. Do one of the
following:

 Press the CLR key to exit the SYS mode and

return to the Setup menu.

 View the results from the last calibration

performed on the selected tag. (Refer to the
section entitled View the Calibration Results
for a Specific Tag
earlier in this section.)

 Select another tag.