Doc. Cal. Mode

7-6

you must return to the Tag Status screen to
specify a different tag.

The Tag Status screen is divided into six sections.
These sections are described below in the order that
they appear on the screen from top to bottom:

Tag Number/Area. The current tag number and area is
listed in the top portion of the screen.

NOTE: When using any instrument maintenance

software packages, all tags may be assigned to
an area. Refer to your instrument maintenance
software manual for more information.

Input/Output Types. These two lines list the field
instrument output type (mA in the example) and the
field instrument input type (PSI in the example).

Calibration Status. This line indicates whether the tag
has been calibrated, and if so, if it passed or failed.

Time/Date. These two lines indicate the time and date
the calibration (if completed) was performed. If the test
has not been performed, the lines read 00:00:00 (as
shown in the example).

Error/Alert Tolerances. These two lines indicate the
error and alert tolerances specified in the tag.

Next. Indicates the next test to be performed.