Martel Electronics BetaGauge 3II User Manual

Page 159

Advertising
background image

Doc. Cal. Mode

7-18

2.

After you select the Zero or Ready menu item for
channel 1, the calibrator prompts you to zero
channel 2. Repeat step 1 to zero the display for
channel 2 or bypass the option.

3.

After you select the Zero or Ready menu item in
step 2, the calibrator displays the Execute screen.
Refer to the paragraphs below for information
concerning the Execute screen.

NOTE: You can exit the SYS mode by pressing
the CLR key once to return to the Tag Status
screen, then once again to return to the Setup menu.

Accessing the Execute Screen

After you specify whether to zero the display readings
for channel 1 and channel 2 on the Zero screen, the
calibrator displays the Execute screen. A sample is
shown below:

The Execute screen is described on the following page.

Tag 2:
BG2 TAGS Tag 2

Inst. OUT: mA
Inst. IN:

PSI

Pt.

IN

OUT

Un 1

3.0000

4.0000

Un 2

6.0000

8.0000

Un 3

9.0000

12.000

Actual

IN

OUT

1

0.0983

-0.0005

Appr. Input:

3.0000

Set & press ENTER