Martel Electronics BetaGauge 3II User Manual

Page 155

Advertising
background image

Doc. Cal. Mode

7-14

The Test Setup screen is divided into five sections.
These sections are described below in the order that
they appear on the screen from top to bottom:

Tag Number/Area. The current tag number and area is
listed in the top portion of the screen.

Low/High Test Point Values. This section lists the
lowest and highest test point values of the selected tag.
These values include the minimum to maximum ranges
for the field instrument’s input and output
measurements. Directly to the right of these values are
the BetaGauge II module assignments (Channel
1/Channel 2) for the field instrument’s input and output
measurements.

Calibration Intervals. This section shows the
calibration interval for each module (Channel 1 and
Channel 2).

Status Message. This section indicates whether the
currently installed modules can accommodate the
assigned test point values for the field instrument’s
input and output measurements. If the appropriate
modules are installed, the status reads Modules Correct.
If one or more inappropriate modules are installed, the
status reads Wrong Module(s).

Menu Item. The Ready menu item appears at the
bottom of the screen only when the status message
reads: Modules Correct. You can use the Ready menu
item to proceed to the Zero screen where you can zero
the display readings for the currently installed modules.

Advertising