Vaisala HMP240 User Manual
Page 100
Advertising
![background image](/manuals/408448/100/background.png)
HMP240
Appendix 2: RS 485/422 serial port module
M210300en
98
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.
•
Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
•
Disconnect the terminal.
•
Check that the jumpers in connector X4 are in the right place
X4
X4
X1
X2
•
Remount the RS 485/422 serial module.
•
Close the cover.
•
Repeat this setting procedure with each transmitter.
•
When all transmitters on the network have been configured, switch
them off.