Vaisala HMP240 User Manual

Page 53

Advertising
background image

HMP240 series

M210300en

User's Guide

51

6.2.1 One point offset calibration

6.2.1.1

With serial commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

After giving the correct temperature value (Ref1) and
pressing <cr>, press any key and then <cr>.

6.2.1.2

With display / keypad commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

Select T 1 point cal (or Ta 1 point cal), change the
reading to correspond to the reference and press ENT.

6.2.1.3

With LED commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

After adjusting the offset point and pressing ENT, the sec-
ond LED from left flashes. Press ENT without changing
the output value.