Appendix 1: Serial commands

M210300en

86

4.

OTHERS

ITEST

Testing the analogue outputs

ITEST <cr>

or

ITEST a b <cr>

a

=

current/voltage of channel 1

b

=

current/voltage of channel 2

The operation of the analogue outputs can be tested by forcing the outputs to
given values which can then be measured with a current/voltage meter from
the analogue outputs. The response to ITEST command gives six out-
puts/parameters. Only the first two are relevant; they show the channel current
or voltage in mA or V. The other four figures contain information for service
purposes only.

Examples:

reading the channel outputs and parameters

>itest <cr>
1.9438 2.3483 1.00694 10.64634 1.97374 2.17665
>

forcing outputs 0.5 V and 4 V to channels 1 and 2

>itest 0.5 4 <cr>
0.5000 4.0000 1.00694 10.62970 1.23336 3.01722
>

releasing the forced control and reading the outputs

>itest <cr>
1.9427 2.3392 1.00731 10.62428 1.97157 2.16978
>

FILT

Setting the averaging time

Disconnect the security lock jumper!

FILT nnnn <cr>

nnnn

=

averaging time (0 - 1024 seconds)

This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.