Vaisala HMP240 User Manual

Page 29

Advertising
background image

HMP240 series

M210300en

User's Guide

27

5.3.4 Analogue output commands

5.3.4.1

Selecting the output (mA/V)

Disconnect the security lock jumper!

Select Mode in the main menu and Analog outputs in the Mode
menu:

Select Mode ( mA / V ). The current settings for channel 1 are dis-
played:

If the settings are correct, press ENT.

If the settings need to be changed, press CL:

the quantity (mA/V) starts flashing; it can be changed with the ar-
row keys and acknowledged with the ENT key

the lower limit starts flashing

acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given one character at a time
with the arrow keys

the upper limit starts flashing

acknowledge the upper limit with ENT or start changing it by pressing
CL; a new upper limit is given one character at a time with the arrow
keys

When channel 1 has been set, the programme goes on to channel 2; the proce-
dure is the same as with channel 1.

NOTE

The analogue output jumpers must also be set to right
places (see Figure 4.2).

Advertising