HMP240 series

M210300en

User's Guide

61

the external load. The output current can, however, be measured at test points
CH1+/CH1- and CH2+/CH2- without disconnecting the output wires. These
test points can therefore be used in one point offset correction against an accu-
rate reference or in checking the current output without disconnecting the
analogue output from the process.

X2

OPENED COVER OF THE HMP 243

X15

CH2

CH1 +

+

C

H

1

+

C

H

1

-

C

H

2

+

C

H

2

-

Figure 7.5

Location of the CH1 and CH2 test points

CH1+

CH1-

TEST POINTS

CH1+

CH1-

mA

R

L

X2

Figure 7.6

Circuit diagram of the analogue output current test points

7.7

Adjusting the contrast of the display

The contrast of the display can be adjusted using the trimmer "LCD display
contrast" located next to press switches.