HMP240
Appendix 1: Serial commands

M210300en

84

When asking the current address, no address number is given:

>ADDR <cr>
Address : 2 ? <cr>

FROST

Setting the calculation mode

Disconnect the security lock jumper!

FROST ON/OFF <cr>

This command is used to select whether the transmitter calculates the
frostpoint (default) or the dewpoint reading at dewpoint temperatures below
0

°

C. Select FROST ON for frostpoint and FROST OFF for dewpoint

calculations. For example:

>Frost <cr>
Frost :

ON

>Frost off

Frost : OFF

>Frost on
Frost : on
>

RESET

Resetting the transmitter

RESET <cr>

Resets the transmitter. All settings that have been changed remain in the mem-
ory even after reset or power failure.

3.1

Operation modes

SMODE

Setting the serial interface

SMODE xxxx<cr>

xxxx

=

STOP, RUN or POLL

In STOP mode: measurements output only by command, all commands can be

used

In RUN mode:

outputting automatically, only command S can be used

In POLL mode: measurements output only with command SEND. When in

POLL mode, the output state is changed as follows:

OPEN aa <cr>