Single loop wiring

Set the address of the transmitter; you can choose any number
between 1 and 99. In this example the address is 22:

>addr 22
Address : 22

Set the serial bus settings according to your network specifications.
This setting will become valid after next RESET or power off:

>seri 2400 e 7 1 h
2400 E 7 1 HDX

Set the transmitter in POLL mode:

>smode poll
Serial mode : POLL

NOTE 1

The SMODE command must be given last.

NOTE 2

The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.

Check that the transmitter responds to its address:

send 22
RH= 24.4 %RH T= 29.1 'C

Disconnect the terminal.

Remount the digital current loop module.

Close the cover.

When all transmitters on the network have been configured, switch
them off.

Dual loop operation

Single pair/dual pair use is configured through wiring (see figure on page 2).

The following procedure must be repeated with all transmitters.

Open the transmitter cover.

Pull out the digital current loop module, if it is already mounted.