Poll mode, Current loop settings – Vaisala HMP240 User Manual

Page 110

Advertising
background image

HMP240
Appendix 3: Digital current loop module

M210300en

108

Ch2 ( T )

hi

180.000 'C

Transducer

:

PRB serial nr : 0
Calibr. date

: 0

When the necessary settings have been given, close the line to transmitters
(the command closes all open lines):

>close
line closed

CLOSE command is always given without address. If no line is open, there
will be no response to the CLOSE commands.

POLL mode

If a transmitter has been set to POLL mode, it will respond only to commands
sent with its address:

send 22
RH= 24.4 %RH T= 29.1 'C

Addresses from 1 to 99 can be used. According to the 20 mA current loop
standard current flows with no transmission on line. A maximum of 6 devices
can be connected on same single loop line, but the number can be increased to
9 by using dual loop wiring.

More than one baud rate can be used on a current loop line. E.g. the host of
the line can use different baud rates with different devices (this makes it pos-
sible to use the same address for several devices).

Current loop settings

HMP240 settings

Single pair

Dual pair

Full duplex/half duplex

HDX

FDX

Echo on/off

OFF

ON

Terminal settings

Single pair

Dual pair

Line feed after carriage return

yes

no

HDX/FDX

FDX

FDX

When terminal is set to general <lf> (line feed) after <cr> (carriage return),
the listings will have two line feeds where also the HMP240 transmitters send
line feed.